Route:close <Chn; Route:open? <Chn; Route:scan; Route:scan <Chn List - Fluke 1529 User Manual

Chub-e4 thermometer readout
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
User's Guide
8.4.3.2
8.4.3.3
8.4.3.4
8.4.3.5
8.4.3.6
8.4.3.7
92
ROUTe:CLOSe <chn>
Enables the specified channel.
ROUT:CLOS 3
The <chn> parameter specifies the channel (1 to 4) to enable. This comand
does not apply to memory channels.
ROUTe:OPEN? <chn>
Returns the state of the specified channel.
Example: ROUT:OPEN? 1
Response: 0
The <chn> parameter specifies the channel (1 to 4). If the specified channel is
enabled this query returns 0. If the specified channel is disabled, this query re-
turns 1.
ROUTe:OPEN <chn>
Disables the specified channel.
Example: ROUT:OPEN 4
The <chn> parameter specifies the channel (1 to 4) to disable. This command
does not apply to memory channels.

ROUTe:SCAN?

Returns a comma delimited list of the enabled channels.
ROUT:SCAN?
(@2,3,4)
The list of channels is enclosed in parenthesis and preceded with the @
symbol.
ROUTe:SCAN <chn list>
Sets the channels for scanning.
ROUT:SCAN 2,3
The <chn list> parameter is a comma delimited list of the channels (1 to 4) to
enable. Channels that are omitted from this list are disabled. If the <chn list>
parameter is omitted, all channels are disabled. The *RST command sets chan-
nel 1 ON, and all other channels OFF.

ROUTe:SCAN:MODE? [MIN|MAX|DEF]

Returns the current scan mode setting.

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents