Channel Mode; Moving Average; Figure 10 Channel Mode - Fluke 1529 User Manual

Chub-e4 thermometer readout
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

hold the Exit button to cancel to the main display or press the EXIT button to
cancel and to move to the next parameter.
7.1.3

Channel Mode

The CHANNEL MODE function sets the channel display mode. When this
function is selected, the bottom portion of the display shows the current scan
mode and allows the setting to be changed.

Figure 10 Channel Mode

The following scan modes are available:
SCAN - measures enabled channels scanning through and displaying the
channels one at a time, sequentially, in large digit format.
SIMULT - measures and displays enabled channels simultaneously.
The LR buttons are used to select a setting. Press the Enter button to save
the setting and exit. Press the Exit button to cancel and to exit to the menu.
If the measure period is 0.1 or 0.2 seconds, the CHANNEL MODE function is
unavailable. The channel mode is automatically set to scan mode. The follow-
ing message is displayed when the CHANNEL MODE function is selected.
NOT AVAILABLE AT THIS RATE.
Press ENTER to continue...
If the measure period is 0.5 seconds or greater, the channel mode operates
normally.
7.1.4

Moving Average

The MOVING AVERAGE function sets the input averaging. The moving aver-
age filter is useful for smoothing variations in the measurements and for im-
24.159C
1
24.142C
3
SC MODE:
SIMULT
1529
CHUB E-4 THERMOMETER READOUT
ENTER / MENU
UNIT
7 Menu Functions
24.634C
2
24.015C
4
EXIT