Calibrate<Chn>:Date:due (<Year>,<Month>,<Day>)|Min|Max|Def; Calibrate<Chn>:Parameter:offset<N>? [Min|Max|Def] - Fluke 1529 User Manual

Chub-e4 thermometer readout
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
User's Guide
8.4.5.5
8.4.5.6
98
CAL1:DATE:DUE?
2001,9,21
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. The response is returned in the format,
<year>,<month>,<day>. Specifying the MIN parameter returns a value of
2000,1,1. Specifying the MAX parameter returns a value of 2099,12,31. Spec-
ifying the DEF parameter returns a value of 2000,1,1.
CALibrate<chn>:DATE:DUE (<year>,<month>,<day>)|MIN|MAX|DEF
Sets the recalibration date for the specified channel.
CAL1:DATE:DUE 2001,8,28
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. The <year>, <month>, and <day> parame-
ters must all be supplied. The <year> parameter is a four-digit number (2000 to
2099). The <month> parameter is a one or two-digit number (1 to 12). The
<day> parameter is a one or two-digit number (1 to 31). Specifying the MIN,
MAX or DEF parameter sets the recalibration date to the minimum, maximum,
or default value respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <pass> command
(see Section 9.4.11, Password Commands).
CALibrate<chn>:PARameter:OFFSet<n>? [MIN|MAX|DEF]
Returns the calibration offset parameter for the specified channel.
CAL1:PAR:OFFS2?
–1.2
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion offset suffix <n> specifies the resistance range parameter for PRTs (1) or
Thermistors (2). For Thermocouple channels, the calibration offset suffix <n>
is ignored and may be omitted. Note: For PRT/Thermistor channels, the value
returned by this command is always in ohms.
For PRT/Thermistor channels, specifying the MIN parameter returns a value of
-9 when the offset suffix <n> is 1 (for PRTs) and -9000 when the offset suffix
<n> is 2 (for Thermistors). For Thermocouple channels, specifying the MIN
parameter returns a value of -9. For PRT/Thermistor channels, specifying the
MAX parameter returns a value of 9 when the offset suffix <n> is 1 (for PRTs)
and 9000 when the offset suffix <n> is 2 (for Thermistors). For Thermocouple
channels, specifying the MAX parameter returns a value of 9. Specifying the
DEF parameter returns a value of 0.

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents