Logging:label<N>:Name <Label; System Commands; Idn; Opt - Fluke 1529 User Manual

Chub-e4 thermometer readout
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

The label suffix, <n>, specifies the number of the label (1 to 25) to retrieve.
The label name is returned in string format.
8.4.7.23
LOGging:LABel<n>:NAME <label>
Sets the name of the specified label.
LOG:LAB18:NAME BATH
The label suffix, <n>, specifies the number of the label (1 to 25) to set. Label
names can be up to 8 characters in length and can include any letters, numeric
digits, and the underscore ('_').
8.4.8

System Commands

The system commands are used for querying instrument identification parame-
ters, temperature units, and for setting operating parameters to the defined de-
fault values.
8.4.8.1

*IDN?

Returns the instrument identification string that indicates the manufacturer,
model number, serial number, and code firmware version.
*IDN?
HART,1529,A09001,1.11
8.4.8.2

*OPT?

Returns a list of input modules installed.
*OPT?
PRT,TC
Input module types are reported by position, top to bottom, of the two input
modules. The returned example is for a 1529 equipped with one 2-channel PRT
input module and one 2-channel Thermocouple input module.
8.4.8.3

*RST

Sets the instrument operating parameters to defined conditions.
*RST
The following parameters are set when this command is issued.
Sets the scanning sequence timer to 1 second.
Sets channel 1 ON and all others OFF.
Sets the scan mode to SCAN.
Sets the moving average to 1.
Sets the printing to all ports OFF.
8 Digital Communications Interface
Commands
109