bit 4: measurement invalid
bits 5-15: not used
8.4.12.18

STATus:QUEStionable:CONDition?

Returns the Questionable Status Condition Register.
STAT:QUES:COND?
16
The assignment of the bits in this register is as follows:
bits 0-3: not used
bit 4: measurement invalid
bits 5-15: not used
8.4.12.19

STATus:QUEStionable:ENABle? [MIN|MAX|DEF]

Returns the Questionable Status Enable Register.
STAT:QUES:ENAB?
255
8.4.12.20
STATus:QUEStionable:ENABle <num>|MIN|MAX|DEF
Sets the Questionable Status Enable Register.
STAT:QUES:ENAB 10