Calibrate<Chn>:Parameter:offset<N> <Num>|Min|Max|Def; Calibrate<Chn>:Parameter:scale<N>? [Min|Max|Def]; Calibrate<Chn>:Parameter:scale<N> <Num>|Min|Max|Def - Fluke 1529 User Manual

Chub-e4 thermometer readout
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

8.4.5.7
CALibrate<chn>:PARameter:OFFSet<n> <num>|MIN|MAX|DEF
Sets the calibration offset parameter for the specified channel.
CAL1:PAR:OFFS2 5
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion offset suffix <n> specifies the resistance range parameter (in ohms) for
PRTs (1) or Thermistors (2). For Thermocouple channels, the calibration offset
suffix <n> is ignored and may be omitted. The <num> paramter specifies the
value for the specified offset parameter. Specifying the MIN, MAX or DEF pa-
rameter sets the offset parameter to the minimum, maximum, or default value
respectively.
This command is password-protected. To access this command, the password
must be successfully entered using the SYST:PASS:CEN <password> com-
mand (see Section 9.4.11, Password Commands).
8.4.5.8
CALibrate<chn>:PARameter:SCALe<n>? [MIN|MAX|DEF]
Returns the calibration scale parameter for the specified channel.
CAL1:PAR:SCAL2?
0
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion scale suffix <n> specifies the resistance range parameter for PRTs (1) or
Thermistors (2). For Thermocouple channels, the calibration scale suffix <n> is
ignored and may be omitted. Note: For PRT/Thermistor channels, the value re-
turned by this command is always in ohms.
For PRT/Thermistor channels, specifying the MIN parameter returns a value of
-9 when the scale suffix <n> is 1 (for PRTs) and -9000 when the scale suffix
<n> is 2 (for Thermistors). For Thermocouple channels, specifying the MIN
parameter returns a value of -9. For PRT/Thermistor channels, specifying the
MAX parameter returns a value of 9 when the scale suffix <n> is 1 (for PRTs)
and 9000 when the scale suffix <n> is 2 (for Thermistors). For Thermocouple
channels, specifying the MAX parameter returns a value of 9. Specifying the
DEF parameter returns a value of 0.
8.4.5.9
CALibrate<chn>:PARameter:SCALe<n> <num>|MIN|MAX|DEF
Sets the calibration scale parameter for the specified channel.
CAL1:PAR:SCAL2 2
The CALibrate suffix, <chn>, specifies the channel (1 to 4). This command
does not apply to memory channels. For PRT/Thermistor channels, the calibra-
tion scale suffix <n> specifies the resistance range parameter (in ohms) for
PRTs (1) or Thermistors (2). For Thermocouple channels, the calibration scale
suffix <n> is ignored and may be omitted. The <num> parameter specifies the
8 Digital Communications Interface
Commands
99

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading

Table of Contents