Logging:automatic:status <Bool; Logging:automatic:time? [Min|Max|Def]; Logging:automatic:time <Num>|Min|Max|Def; Logging:automatic:value? <Num>|Min|Max|Def - Fluke 1529 User Manual

Chub-e4 thermometer readout
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
User's Guide
8.4.7.10
8.4.7.11
8.4.7.12
8.4.7.13
106
0
If this query returns a 0, an auto log session is not running. If this query returns
a 1, an auto log session is running.
LOGging:AUTomatic:STATus <bool>
Sets the auto log status.
LOG:AUT:STAT 0
The <bool> parameter turns auto logging on (1 or ON) or off (0 or OFF). If si-
multaneous scan mode is enabled, one measurement is stored for each channel
in the auto log. The *RST command set the auto log status to off.

LOGging:AUTomatic:TIMe? [MIN|MAX|DEF]

Returns the auto log interval setting.
LOG:AUT:TIM?
1
Specifying the MIN parameter returns a value of 0.1. Specifying the MAX pa-
rameter value returns value of 3600. Specifying the DEF parameter returns a
value of 1. The units are seconds.
LOGging:AUTomatic:TIMe <num>|MIN|MAX|DEF
Sets the auto log interval setting.
LOG:AUT:TIM 1
The <num> parameter can be one of the following values: 0.1, 0.2, 0.5, 1, 2, 5,
10, 30, 60, 120, 300, 600, 1800 or 3600. The units are seconds. If a value other
than one listed is used, the instrument automatically selects the nearest accept-
able value that is less than the value entered. Values greater than 3600 or less
than 0.1 are ignored. Specifying the MIN, MAX or DEF parameter sets the log
interval setting to the minimum, maximum, or default value respectively.
LOGging:AUTomatic:VALue? <num>|MIN|MAX|DEF
Returns the specified auto log entry.
LOG:AUT:VAL? 10
,1,22.860,C,12,19,46,,,
The <num> parameter specifies the auto log entry to return. Specifying either
the MIN or DEF parameter returns the first entry in the auto log. Specifying the
MAX parameter returns the last entry in the auto log.
Note: Every time an auto log session is started, the first entry in the log for that
session consists of the data label and the date. All subsequent entries in the auto
log for that session consist of the channel, measurement value, units and time.