System:boot:version; System:code:version; System:error; System:snumber - Fluke 1529 User Manual

Chub-e4 thermometer readout
Hide thumbs Also See for 1529:
Table of Contents

Advertisement

1529 Chub-E4 Thermometer Readout
User's Guide
8.4.8.4
8.4.8.5
8.4.8.6
8.4.8.7
8.4.8.8
110
Sets the units to C.
Sets the resolution to AUTO.
Stops auto logging.
Clears the statistical functions.

SYSTem:BOOT:VERSion?

Returns the boot ROM version.
SYST:BOOT:VERS?
0.50

SYSTem:CODE:VERSion?

Returns the firmware code version.
SYST:CODE:VERS?
1.11

SYSTem:ERRor?

Returns a system error message if any are present in the system error queue.
SYST:ERR?
0,"No error"
Each error condition produces only one error message at a time. The error
queue may contain up to ten messages. The messages are reported in the order
they occured. Reading a message from the queue removes the message so the
next message can be read. If more than ten errors occur before being read, the
tenth error in the queue is -350,"Queue overflow" and all but the first nine er-
rors are discarded. All errors cause the Error Bit (ERR, bit 2) for the Status
BYTE Register to be set. Refer to Troubleshooting section for the error
messages.

SYSTem:SNUMber?

Returns the instrument serial number.
SYST:SNUM?
A09002
If the serial number has not been set, a 0 is returned. The serial number is set at
the factory and cannot be changed!

SYSTem:VERSion?

Returns the SCPI version number.
SYST:VERS?

Hide quick links: