Procedure; Visual Inspection; 1523/24 Calibration Parameters (As Found); Table 29 1523/24 Accuracy Test Settings And Specifications - Voltage - Fluke 1523 Technical Manual

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

1523, 1524 Reference Thermometer
Manual Calibration Process
7.9.1
7.9.1.1
7.9.1.2
82
1.800.544.2843
Resistance
Reference
Range
(Nominal) (Ω)
HI_OHMS
40k
100k
300k
500k
RJ_OHMS
10k
Table 29 1523/24 Accuracy Test Settings and Specifications – Voltage
Reference
Soak
(Nominal) (mV)
(Sec)
-10
20
0
20
25
20
75
20

Procedure

Visual Inspection

1523/24 Calibration Parameters (As Found)

NOTE: If not taking As Found data, skip to the 1523/24 alignment section.

Table 30 As Found Readout/Calibration Parameter Settings

Parameter
Description
# of Cal Ref Points
Each Cal Range has 1 to 4 Ref Cal Points
Cal Ref Points
Reference values set to each cal point
Cal Adj values
Offsets used at each Cal Ref point
1. Using a terminal program set the terminal program communication settings to:
Baud
9600
Parity
None
Flow
Control None
Stop Bits
1
www.calcert.com
15
10
20
5
25
0
30
Soak (Sec)
Samples
20
40
20
40
20
40
20
40
20
40
StdDev
Spec
Spec
Samples
(±mV)
(± mV)
40
0.0055
0.00138
40
0.005
0.00125
40
0.0063
0.00156
40
0.0088
0.00219
Spec
Std Dev
(±Ω)
Spec (±Ω)
12.0
3.0
30.0
7.5
90.0
22.5
150.0
37.5
5.0
1.25