50S & 50D
Users Manual
Changing to the J-Type Thermocouple
Press HOLD during power-up.
Changing to Low Resolution
To select the low display resolution [1.0
1.
Turn the thermometer off.
Press T1 + ON/OFF for 2 to 3 seconds.
2.
The alternate resolution remains set until the
thermometer is turned off.

Adjusting the Offset

To compensate for thermocouple errors:
1.
Plug the thermocouple into the input connector.
2.
Place the thermocouple in a known, stable temperature
environment (such as an ice bath or a dry well
calibrator) and allow the readings to stabilize.
3.
Simultaneously push and turn the OFFSET control until
the display shows the correct reading.
This adjustment affects calibration.
8
o
o
C (1.0
F)]: