Procedure; Visual Inspection; 1523/24 Calibration Parameters (As Found); Table 29 1523/24 Accuracy Test Settings And Specifications - Voltage - Fluke 1523 Technical Manual

Reference thermometer
Hide thumbs Also See for 1523:
Table of Contents

Advertisement

1523, 1524 Reference Thermometer
Manual Calibration Process
Resistance
Range
HI_OHMS
RJ_OHMS
Table 29 1523/24 Accuracy Test Settings and Specifications – Voltage
(Nominal) (mV)
7.9.1

Procedure

7.9.1.1

Visual Inspection

7.9.1.2

1523/24 Calibration Parameters (As Found)

NOTE: If not taking As Found data, skip to the 1523/24 alignment section.

Table 30 As Found Readout/Calibration Parameter Settings

Parameter
# of Cal Ref Points
Cal Ref Points
Cal Adj values
1. Using a terminal program set the terminal program communication settings to:
82
Reference
(Nominal) (Ω)
40k
100k
300k
500k
10k
Reference
Soak
(Sec)
-10
20
0
20
25
20
75
20
Description
Each Cal Range has 1 to 4 Ref Cal Points
Reference values set to each cal point
Offsets used at each Cal Ref point
Baud
9600
Parity
None
Flow
Control None
Stop Bits
1
Soak (Sec)
Samples
20
40
20
40
20
40
20
40
20
40
Spec
Samples
(±mV)
40
0.0055
40
0.005
40
0.0063
40
0.0088
Spec
Std Dev
(±Ω)
Spec (±Ω)
12.0
3.0
30.0
7.5
90.0
22.5
150.0
37.5
5.0
1.25
StdDev
Spec
(± mV)
0.00138
0.00125
0.00156
0.00219