Triggering on Edges

If the signal is unstable or has a very low frequency, use
edge triggering to obtain full manual trigger control.
To trigger on rising edges of the input A waveform, do the
following:
1
Display the
TRIGGER
2
Open the
F4
3
Open the
ENTER
key labels.
TRIGGER
menu.
TRIGGER OPTIONS
menu.
TRIGGER ON EDGE
When Free Run is selected, the test tool updates the
screen even if there are no triggers. A trace always
appears on the screen.
When On Trigger is selected, the test tool needs a trigger
to display a waveform. Use this mode if you want to
update the screen only when valid triggers occur.
When Single Shot is selected, the test tool waits for a
trigger. After receiving a trigger, the waveform is displayed
and the instrument is set to HOLD.
In most cases it is advised to use the Free Run mode:
4
Select Free Run, jump to Trigger
Filter.
ENTER
5
Set Trigger Filter to Off.
ENTER
Observe that the key labels at the bottom of the screen
have adapted to allow further selection of specific edge
trigger settings:
Triggering on Waveforms
Triggering on Edges