For mVs select probe type 'Voltage'.
For mAs select probe type 'Current'.
For mWs select mathematical function x,
and probe type 'Voltage' for one channel
and 'Current' for the other channel.
Using Cursors on a Mathematical Result (+ - x)
Waveform
Cursor measurements on, for examle, a AxB waveform
give a reading in Watts if input A measures (milli)Volts and
input B measures (milli)Amperes.
For other cursor measurements on, for example, a A+B,
A-B or AxB waveform no reading will be available if the
input A and input B measurement unit are different.
Notes
Using Replay, Zoom and Cursors
Making Cursor Measurements
Using Cursors on Spectrum Measurements
To do a cursor measurent on a spectrum, do the following:
1
From Spectrum measurement
CURSOR
display the cursor key label.
Move the cursor and observe the
2
readings at the top of the screen.
3