To set the Power Meter Preferences:
1. Push
2. In the Power Meter readout screen, push the Sensor 1 or Sensor 2 softkey to
3. Push the Sensor Preferences softkey to show the Power Meter Preferences
4. Sequentially select each of the preference fields with the softkeys to the right
5. While each field is selected, use the softkeys along the bottom of the display
6. To exit the screen, push the Back softkey, or push one of the function keys

Power Reading Averaging and Triggering

The power readout averaging and trigger preferences settings and corresponding
GPIB commands control the averaging factor and type applied by the power
sensor when averaging readings for display or output via GPIB.
When the Averaging preference is set to Auto, the power sensor continuously
determines the averaging factor depending on the power level with a maximum
settling time of 4 seconds for the sensor's averaging filter. Alternatively, a specific
averaging factor value between 1 and 32768 in a 2
When the Trigger preference is set to Continuous, the power readout shows the
rolling average of the power sensor readings.
3-100
to show the Power Meter readout screen as shown in Figure 3-39.
select the required sensor channel. The focus, indicated by the highlight box
around the sensor reading field, and the input channel number within the
Sensor Preferences softkey legend, will change according to the selection
made. The selection and will remain until the other sensor is subsequently
selected.
screen shown in Table 3-32.
of the screen.
or the spin wheel to choose a preference.
(, , , or ) or .
n
sequence may be selected.