Fluke 190 Series II User Manual page 108

Scopemeter 190 series ii
Hide thumbs Also See for 190 Series II:
Table of Contents

Advertisement

Fluke 190 Series II
Users Manual
A message appears asking you whether to start the 10:1
probe calibration.
6
Start the probe calibration.
F4
A message appears telling you how to connect the probe.
Connect the red 10:1 voltage probe to input A and to the
probe calibration reference signal as shown in Figure 51.
7
Adjust the trimmer screw in the
probe housing until a pure square
wave is displayed.
For instructions to access the
trimmer screw in the probe housing
see the probe instruction sheet.
8
Continue with DC calibration.
F4
Automatic DC calibration is only
possible for 10:1 voltage probes.
100
The test tool automatically calibrates itself to the probe.
During calibration you should not touch the probe. A
message indicates when the DC calibration has completed
successfully.
9
Return.
F4
Repeat the procedure for the blue 10:1 voltage probe on
input B, the gray 10:1 voltage probe on input C and the
green 10:1 voltage probe on input D.
When using 100:1 voltage probes, choose 100:1
attenuation to perform an adjustment.
Note