Appendix B Verification and Calibration

Verification

Performance
Calibration
Equipment Required
Type
Digital Voltmeter
Current Monitor
Load Resistor
Electronic Load
GPIB Controller
Oscilloscope
RMS Voltmeter
Differential Amplifier
Terminations
Variable-voltage xfmr
or AC source
94
Verification procedures verify that the power supply is operating
normally and is within published specifications. There are two types
of verification tests:
These tests verify that the power supply meets all of the specifications listed
in Appendix A. They can also be used to verify that the power supply is
properly calibrated.
These procedures calibrate the power supply.
If the power supply fails any of the tests or if abnormal test results
are obtained, try calibrating the unit. If calibration is unsuccessful,
return the unit to an Agilent Technologies repair facility (see
Appendix D).
The equipment listed in the following table, or the equivalent to this
equipment, is required for the calibration and performance tests. A
test record sheet may be found at the back of this section.
Specifications
Resolution: 10 nV @ 1V;
Readout: 8 1/2 digits
Accuracy: 20 ppm
15 A (0.1Ω) 0.04%, TC=5ppm/°C
100 A (0.01Ω) 0.04%, TC=5ppm/°C
300 A (0.001Ω) 0.04%, TC=5ppm/°C
For 750 W models:
0.06Ω, 0.09Ω, 0.21Ω, 0.53Ω, 1.20Ω, 2.11Ω, 4.80Ω, 8.42Ω,
13.33Ω, 30.0Ω, 120Ω, 462Ω - all resistors 1kW minimum.
For 1500 W models:
0.03Ω, 0.04Ω, 0.10Ω, 0.26Ω, 0.60Ω, 1.05Ω, 2.40Ω, 4.21Ω,
6.67Ω, 15.0Ω, 60Ω, 231Ω - all resistors 2kW minimum.
150 V, 100 A minimum (for Models N5741- N5750A)
150 V, 180 A minimum (for Models N5761- N5770A)
Full GPIB capabilities (for calibrating the unit over the GPIB)
Sensitivity: 1 mV
Bandwidth Limit: 20 MHz
Probe: 1:1 with RF tip
True RMS
Bandwidth: 20 MHz
Sensitivity: 100 µV
Bandwidth: 20 MHz
1 – 50Ω BNC termination
2 – 50Ω, 1/8 W termination resistors
Adjustable to highest rated input voltage range.
Power: 2000 VA
Recommended Model
Agilent 3458A or equivalent
Guildline 9230/15
Guildline 9230/100
Guildline 9230/300
Agilent N3300A mainframe,
with 3 - N3305A modules
Agilent 82350B or
equivalent
Agilent Infiniium or
equivalent
Rhode and Schwartz Model
URE3 or equivalent
LaCroy 1855A, DA1850A,
or equivalent
Agilent 6813B or equivalent
Series N5700 User's Guide