Compatibility Command Summary - Agilent Technologies N5700 Series User Manual

System dc power supply
Hide thumbs Also See for N5700 Series:
Table of Contents

Advertisement

Compatibility Command Summary

Compatibility Command
ASTS?
CLR
DLY <delay>
DLY?
ERR?
FAULT?
FOLD
FOLD?
HOLD
HOLD?
ID?
IMAX
IMAX?
IOUT?
ISET <current>
ISET?
OUT <on|off>
OUT?
Series N5700 User's Guide
The following table documents the compatibility commands that the
Agilent N5700 power supplies support. All compatibility commands
are accepted; however, some commands do nothing.
Description
Queries the accumulated status (ASTS). The response represents the sum
of the binary weights of the ASTS register bits. The ASTS register is set to
the present status after being queried.
Returns the power supply to the power-on state. Same as *RST.
Generates error 203.
Generates error 203.
Queries the present programming or hardware error. An error code number
is returned over the GPIB to identify the error. The error register is cleared
after being read.
Queries the fault register. A bit is set in the fault register when the
corresponding bit in both the status and the mask registers. The response
is an integer 0 to 255. The fault register is cleared after being read.
Turns the OCP on or off. This is only allowed for constant current mode
(FOLD 2). Constant voltage mode (FOLD1) generates error 203.
Queries the OCP setting. The response is FOLD 2.
When turned on (HOLD 1), causes the VSET, ISET, FOLD, and UNMASK
values to be held until a trigger occurs. This only applies to the
compatibility functions, not the SCPI functions
Queries the hold setting. The response is HOLD 1.
Queries the identification (model number) of the power supply.
Sets a soft programming limit for current. Attempting to program the
current above this setting will generate an error.
Queries the IMAX setting. The response is a real number.
Queries the measured output current. The response is a real number.
Sets the output current.
Queries the present current setting. The response is a real number.
Turns the output on or off. On/off equals 1 turns the output on; equals 0
turns the output off.
Queries whether the output is turned on or off . The response is OUT 1 (on)
or OUT O (off). The front panel displays OFF when the output is off.
Compatibility Appendix D
Similar SCPI
Command
STAT:OPER:EVEN?
STAT:QUES:EVEN?
*ESE?
*RST
SYST:ERR?
STAT:OPER?
STAT:QUES?
*ESE?
CURR:PROT:STAT
CURR:PROT:STAT?
VOLT:TRIG
CURR:TRIG
*IDN?
MEAS:CURR?
CURR
CURR?
OUTP:STAT
OUTP:STAT?
125

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents