Agilent Technologies N5700 Series User Manual page 45

System dc power supply
Hide thumbs Also See for N5700 Series:
Table of Contents

Advertisement

PROGRAMMING
OPTIONAL SETS
LOWER LIMIT
External Monitoring of Output Voltage and Current
Series N5700 User's Guide
Set the programming resistors to the desired resistance and
turn the power supply on. Adjust the resistors to change the
power supply output.
The analog control circuits let you set the output voltage and current
limit up to 5% over the model-rated maximum value. The power
supply will operate within the extended range, however it is not
recommended to operate the power supply over its voltage and
current rating, and performance in this region is not guaranteed.
SW1 switch 3
Down (default)
Up
CURRENT LIMIT
PROGRAMMING
RESISTOR
13
25
OPTIONAL SETS
UPPER LIMIT
The J1 connector also provides analog signals for monitoring the
output voltage and current. Selection of the voltage range between 0
– 5 V or 0 – 10 V is made by SW1 setup switch 4. The monitoring
signals represent 0 to 100% of the power supply output voltage and
current rating. The monitor outputs have a 500 Ω series output
resistance. Make sure that the sensing circuit has an input resistance
greater than 500 kΩ or the accuracy will be reduced.
SW1 switch 4
Down (default)
Up
J1 pin 12 is the signal common for J1 pins 11 and 24.
Operating the Power Supply Locally 3
Voltage Programming
(J1 pin 9)
0 – 5 kΩ
0 – 10 kΩ
OUTPUT VOLTAGE
PROGRAMMING
12
10
9
8
23
22
Voltage
J1 signal
range
connection
0 – 5 V
J1 pin 11
J1 pin 24
0 – 10 V
J1 pin 11
J1 pin 24
Current programming
(J1 pin 10)
0 – 5 kΩ
0 – 10 kΩ
PROGRAMMING
RESISTOR
1
14
OPTIONAL SETS
LOWER LIMIT
OPTIONAL SETS
UPPER LIMIT
Signal function
Voltage Monitor
Current Monitor
Voltage Monitor
Current Monitor
45