753/754
Users Manual

Frequency Measurement

Ranges
[1]
1.00 Hz to 110.00 Hz
110.1 Hz to 1100.0 Hz
1.101 kHz to 11.000 kHz
11.01 kHz to 50.00 kHz
Coupling: AC
Minimum Amplitude for Frequency Measurement (square wave):
<1 kHz: 300 mV p-p
1 kHz to 30 kHz: 1.4 V p-p
>30 kHz: 2.8 V p-p
Maximum input:
<1 kHz: 300 V rms
>1 kHz: 30 V rms
Input Impedance: >4 MΩ
[1] For frequency measurement less than 110.00 Hz, specifications apply for signals with a slew rate >5 volt/millisecond.
±

DC Voltage Output

Range
±100.000 mV
1 μV
±1.00000 V
10 μV
±15.0000 V
100 μV
Maximum Output Current: 10 mA, In the 100 mV range add 0.010 mV to specification when sourcing >1 mA.
For sourcing dc voltages <110.000 mV, accuracy is not specified in RF fields >1 V/m, 80 MHz to 700 MHz.
Temperature Coefficient: 0.001 % of output + 0.001 % of range / °C (<18 °C or >28 °C)
110
Find Quality Products Online at:
Resolution
0.01 Hz
0.1 Hz
0.001 kHz
0.01 kHz
Resolution
1 Year
0.01 % + 0.005 mV
0.01 % + 0.00005 V
0.01 % + 0.0005 V
GlobalTestSupply
www.
2 Year
0.05 Hz
0.5 Hz
0.005 kHz
0.05 kHz
% of Output + Floor
2 Year
0.015 % + 0.005 mV
0.015 % + 0.00005 V
0.015 % + 0.0005 V
.com
sales@GlobalTestSupply.com

Advertisement

Table of Contents
loading

This manual is also suitable for:

754

Table of Contents