Initialize New Batteries

When new batteries are installed, the internal coulomb counter needs to be
reset to ensure accurate data on the battery's remaining life.
From the logger options menu, select "Initialize new batteries."
A warning will appear asking the user if they wish to initialize new batteries.
Select "Yes".
Press the battery type on the menu in the appearing dialog box. This opens a
panel where the user can scroll between compatible battery options and choose
the appropriate battery type installed.
When the battery is appropriately chosen, press the "Select" icon. After a
loading period, a message will appear confirming the successful initialization of
the battery.
PSM0005A
Released: 23 Jan 2025
RST Instruments Ltd.
RSTAR Affinity Tilt Logger
Installation and User Manual