Download Print this page

Environmental Conditions; Storage Instructions; Cleaning Instructions; Related Manuals - Getinge Applikon my-Control Hardware Manual

For minibio reactors 250, 500, 1000 ml

Advertisement

THE my-CONTROL
3.9

ENVIRONMENTAL CONDITIONS

The my-Control may be used at locations with the following environmental conditions:
Use
Altitude:
Temperature:
Relative humidity:
Mains supply voltage:
Transient over voltages
Pollution:
Power consumption:

3.10 STORAGE INSTRUCTIONS

When the my-Control, after being used, has to be stored for a longer time (> 1 month), follow the
instructions below:
• Disconnect the sensor cables and stirrer motor cable from the my-Control. Store the sensors according to
the instructions in the User Manual that comes with the sensor. Store the Stirrer Motor in its original box.
• Remove the pump tubing from the Pumps of the my-Control. Drain and rinse the pump tubing and store
it together with the feed bottles.
• Disconnect the gas supply lines from the gas inlets at the rear panel of the my-Control cabinet.
• Purge the gas supply lines (especially the O
Cover the gas inlets and outlets with blind caps.
• (if necessary) Clean the outside of the my-Control cabinet according to the instructions below.
• Pack the my-Control in its original packaging (plastic bag and box).
• Store it at altitude, temperature and humidity as specified in the section on Environmental Conditions.
• Clean and dry the bioreactor with all its auxiliaries and store it, preferably in its original packaging.

3.11 CLEANING INSTRUCTIONS

The my-Control cabinet may be cleaned with a moist tissue in combination with a non-abrasive cleaner.

3.12 RELATED MANUALS

Also refer to the following Manuals:
• Software Reference Manual,
• Operator Manual,
• PID Control Manual,
• USB Sensor Manuals.
18
indoor
up to 2000 m
4 °C to 40 °C
maximum 80 % for temperatures up to 31 °C,
decreasing linearly to 50 % relative humidity at 40 °C
220 – 240 / 100 - 120 VAC, 50/60 Hz
according to Installation category II
Pollution degree 2 in accordance with IEC 664
P-max = 400 VA
and CO
2
lines) inside the my-Control with instrument air.
2
Hardware Manual my-Control MiniBio