• This set of equipment is designed for bioprocess control and must not be used for other pur- poses. • All Livit Links must be used in combination with Getinge controllers and must not be used with other devices (i.e. personal computers, laptops, etc.), with the exception of the pH and dO Livit Links.
EN 61010-1:2010 / EN 61010-2-010:2020 / EN 61010-2-051:2021 2014/30/EU (EMC) EN 61326-1:2021 / EN 61000-3-2:2019 / EN 61000-3-3:2013 *For Livit Flex, the Machinery Directive includes the requirements of the Low Voltage Directive. Digitally signed by Lukas Lukas van Mierlo Delft,...
Page 82
The powerfail software functionality will only function if the powerfail hardware is also installed. If the software is upgraded to version 2.7, then the proper hardware should also be installed for the functionality to work. Contact Getinge Applikon for the hardware installation.
HMI, system with different sensors and actuators. It is possible to reconfigure a system. To reconfigure the system refer to the local Getinge Applikon representative. For further information about bioreactor and system configuration refer to the relevant sections in the Software Part of this manual.
Tap water is normally used as medium for heating and cooling. The use of deionized water will cause corrosion inside the Thermo Circulator. For conditions other than specified above (e.g. use of a chiller with organic solvents), contact Getinge Applikon or your local distributor.
The stirrer motors are virtually maintenance free since they do not contain carbon brushes. However, the ball bearing needs to be replaced every 20,000 operating hours. Please return the stirrer motor to the Technical Support department of Getinge Applikon or its repre- sentative.
The Analog Input/Output Livit Link (left image below) can be used to connect analog devices, ac- tuators/outputs such as external pumps or analog inputs such as analog sensors to the Livit Flex controller. It allows for connection of up to 4 analog inputs and 2 analog outputs. The following specifications apply: •...
Same rights as operator and can additionally Default Engineer 0000 configure the application. Default Service Reserved for Getinge Applikon For service engineer work. 2.2 Login Procedure By pressing on the LOG IN button in the main top bar, the menu with the login levels...
Page 294
Select the File Format and Data Interval from the drop-down menus. An HMI on an instrument, such as a Livit Flex, can only export raw and CSV files. When the application is running on a Windows computer, raw, CSV and Excel files can be exported.
Page 305
In the absence of these two conditions the HMI will no longer be accessible and the IP address will need to be restored. If this happens contact Getinge Applikon for support. The DHCP (Dynamic Host Configuration Protocol) can be enabled to automatically receive an IP address from a DHCP server in the network.
Page 310
The OPC control tag is a timer with a maximum value of 60 seconds. When a value is received by the Livit Flex controller while the timer is active, the user is logged off and locked out for this period of time.
Page 328
The Instrument filter filters the alarms based on the instrument: • All: shows alarms from all instruments • Livit Flex: shows alarms from the Livit Flex instrument This filter is useful when more than one instrument is connected to the HMI and the user wants to check the alarms from only one instrument.
Page 358
Getinge is a global provider of innovative solutions for operating rooms, intensive care units, sterilization departments and for life science companies and institutions. Based on our firsthand experience and close partnerships with clinical experts, healthcare professionals and medtech specialists, we are improving everyday life for people –...
Need help?
Do you have a question about the Livit Flex and is the answer not in the manual?
Questions and answers