Weight voltage value
3.6.1 The weight calibration scale
In the setting page, click "Static calibration" to enter the interface of weight
calibration, follow the steps on the touch screen to calibrate the scale, and click "Exit" to
return to the main interface after the calibration is completed.For specific operation
methods, please refer to "Operation Instructions on Buttons and Operation Frames" in
Section 3.3.2.
➢ During calibration, ensure that the weighing scale is in the stopped state; otherwise,
the interface of the weighing scale cannot be entered;Calibration should ensure that
there is no item on the weighing platform, no vibration on the weighing platform, and
no relatively strong air flow around the weighing scale.
➢ When the weighing platform is empty, ensure that the weighing platform is at zero
position and stable. Otherwise, please eliminate interference and click "Calibrate the
weighing platform". The second step can only be carried out when the touch screen
indicator is 0 and the stability sign is lit.
➢ When placing the weight, try to avoid the weight hitting the surface of the weighing
platform. Enter the correct weight into the weight box of the weighing platform,
otherwise it will lead to inaccurate calibration scale or calibration failure (the weight of
the weighing platform should be greater than the product weight and not exceed the
maximum range of the weighing scale).
➢ If the calibration fails, please check whether the scale is stable and whether the sensor
is interfered by the outside world or whether the scale is in contact with other
equipment, and re-calibrate after troubleshooting.
3.6.2 Dynamic calibration of scale
On the product page, click "Dynamic calibration" to enter the dynamic calibration
interface, and perform dynamic calibration according to the text prompts. When completed,
relevant parameters will be automatically calculated and generated and product parameters
will be written.Click "Exit" to return to the main interface after completion of calibration.For
details, please refer to Section 3.3.2 "Operation Instructions for Buttons and Operation
Frames".
The voltage value of the sensor after placing the weight