General Measure ChexGo CW-60K-V User Manual page 29

Vertical checkweigher (mcgs+c01 version)
Table of Contents

Advertisement

Description of dynamic calibration parameters:
Name
Zeroing
Get the weight
Dynamic calibration
times
Start up
Maximum weight
Average weight
Correction factor
Figure 3-12 Dynamic calibration screen example
Instructions
Clear the current weight value to zero
Stop the belt running and put the test material on after the
static weight value shown
The number of repeated runs of dynamic calibration. The
default is ten, and no less than five are recommended
The belt will run, and after the statically weighed object has
passed, it should be moved back to the front stage, repeatedly
running dynamically, and the controller will automatically
record the dynamic weight. General dynamic calibration ten
times, the number of dynamic calibration to check the weight
scale will automatically stop
The maximum value of the weight check result during dynamic
calibration
During dynamic calibration, the average weight of each check
will be updated after the number of dynamic calibration is
completed
The standard value is 1000. After the number of dynamic
calibration is completed, the controller will automatically