➢ Calibration must ensure that the check scale in the stopped state, otherwise can not
enter the dynamic calibration interface;When calibrating, ensure that there is no item
on the scale, no vibration on the scale, and no relatively strong air flow around the
scale.
➢ When the weighing platform is empty, it should be ensured that the weighing platform
is at zero position and stable, otherwise, please eliminate external interference and
carry out "zero clearing" operation.
➢ When placing the product, the product should avoid hitting the surface of the weighing
platform. Only after the weight is stable can you click "Obtain static weight";If the
product has a gross weight value, please set the gross weight value first before
dynamic calibration.
➢ The default value of learning times is 10. If the learning result is not accurate, you can
increase the learning times appropriately.If the production requirement is not high, the
learning times can be appropriately reduced to improve the learning speed;External
interference should be avoided in the learning process, and the system will
automatically save the learning results and display them after the completion of
learning.
➢ The change of product detection speed requires dynamic calibration again.
➢ The principle of dynamic calibration, attention to implementation and alternative
methods are introduced in detail in the interface of dynamic calibration parameter help.
If necessary, you can click Help to view.
calculate this value according to the dynamic result and static
weight.
If it is not convenient for dynamic calibration, you need to
enter this value manually, which can be set in the product
parameter interface, and fill in the size of the value refer to the
method in the dynamic calibration parameter help interface
-24-