Settings - Xylem wtw MIQ/IDS Operating Manual

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

MIQ/IDS
Making the set-
tings
Setting
Measuring mode
Temperature mode
ORP shift
(only with measuring
mode mV)
Temp. adjustment
Save and quit
Quit
ba77277d01
Check:
Step 2:
1.
2.
Bring the sensor into the measuring position.
Step 3:
Once the check has been completed, the maintenance condition
has to be switched off manually. More detailed information on the
maintenance condition is given in the IQ S
operating manual.
5.3

Settings

Using <S>, switch from the measured value display to the main menu of the
settings. Then navigate to the setting menu of the sensor. The procedure is
described in detail in the IQ S
Selection/Values
U (mV)
°C
°F
-1250 mV ... +1250
mV
-1.5 °C ... +1.5 °C
or
-2.7 °F ... +2.7 °F
-
-
01/2024
When operating the sensor in a flow through armature:
Remove the sensor from the flow armature.
Submerse the sensor in the test solution.
Compare the measured value to the nominal value.
N
ENSOR
ET
Explanation
Unit of the measured value in the measured
value display.
Unit of the temperature measured value (Celsius,
Fahrenheit).
Here you can set the ORP zero point.
The temperature compensation function enables
the temperature sensor to be balanced against a
reference temperature measurement.
Offset of the zero point:
±1.5 °C (±2.7 °F)
Displayed unit of the temperature:
Temperature mode
see
The settings are stored.
The display switches to the next higher level.
The settings are not stored.
The display switches to the next higher level.
ENSOR
system operating manual.
setting
ORP
N
system
ET
25