Calibration Result; Settings; Messages - Xylem wtw MIQ/IDS Operating Manual

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

Chlorine
Possible calibration
results
Calibration history
Making the set-
tings
Setting
Measuring mode
Sensor
Save and quit
Quit
44
After the calibration procedure has been finished the mainte-
Step 3:
nance condition has to be switched off manually. More detailed
information on the maintenance condition is given in the
IQ S
ENSOR

8.2.3 Calibration result

The externally determined value of the chlorine concentration is evaluated
during input together with the measured reference voltage from the
IQ S
N
.
ENSOR
ET
A chlorine concentration outside the permitted limits cannot be entered.
The calibration history contains the data of the active calibration and the chrono-
logical list with the last calibrations.
The calibration history provides the following information:
Date
Date of the calibration
Raw
Raw value (reference voltage in mV)
Concentration entered (mg/L)
Conc.
8.3

Settings

<S>
Using
, switch from the measured value display to the main menu of the
settings. Then navigate to the setting menu of the sensor. The procedure is
described in detail in the IQ S
Selection/Values
Concentration (mg/l)
Concentration (ppm)
FCML 412 M12-2
FCML 412 M12-20
-
-
8.4

Messages

IDS chlorine sensors do not generate any messages.
N
system operating manual.
ET
N
system operating manual.
ENSOR
ET
Explanation
Measured parameter in the measured value dis-
play
Select the connected sensor.
The sensor determines the measuring range.
The settings are stored.
The display switches to the next higher level.
The settings are not stored.
The display switches to the next higher level.
MIQ/IDS
ba77277d01
01/2024