Calibration Result - Xylem wtw MIQ/IDS Operating Manual

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

MIQ/IDS
Calibration
evaluation
Possible calibration
results
ba77277d01
Display
Init calibration
* Rinse and dry the sensor.
* Immerse the sensor at an
angle in standard 1 (0 FNU/
NTU).
* Position the sensor (see oper
ating manual MIQ/IDS).
* Rinse and dry the sensor.
* Immerse the sensor at an
angle in standard 2 (124 FNU/
NTU).
* Position the sensor (see oper
ating manual MIQ/IDS).
* Rinse and dry the sensor.
* Immerse the sensor at an
angle in standard 3 (1010 FNU/
NTU).
* Position the sensor (see oper
ating manual MIQ/IDS).
Storing calibration
After the calibration procedure has been finished the mainte-
Step 3:
nance condition has to be switched off manually. More detailed
information on the maintenance condition is given in the
IQ S

9.2.4 Calibration result

After calibrating, the calibration data and current state of the sensor are evalu-
ated automatically.
A calibration procedure can have the following results:
Display after calibrating
Measured value display
01/2024
N
system operating manual.
ENSOR
ET
Log book entries
(meaning/actions)
Sensor was successfully calibrated.
Calibration data see calibration history.
Explanation
Confirm with <OK>.
Measuring standard
As soon as a stable measured
value is detected, the next
display appears.
-
Confirm with <OK>.
Measuring standard
As soon as a stable measured
value is detected, the next
display appears.
-
Complete the calibration as a
2-point calibration with <M>.
or
Confirm with <OK>.
Measuring standard
-
As soon as a stable measured
value is detected, the next
display appears.
Confirm with <OK>.
The display returns to the
measured value display. Cali-
bration is finished.
Turbidity
.
.
.
49