Table of Contents

Advertisement

ORP

Prerequisites
24
5
ORP
The MIQ/IDS module connects IDS sensors to the IQ S
Suitable sensors: see section 5.5 A
5.1

Measuring

CAUTION
Contact with the sample can lead to danger to the user!
Depending on the type of sample, suitable protective mea-
sures must be taken (protective clothing, protective goggles,
etc.).
ORP sensors are not calibrated. You can, however, check ORP
sensors by measuring the ORP of a test solution and comparing the
value with the nominal value (see section 5.2).
 The IQ S
N
ENSOR
(see IQ S
N
ENSOR
 The sensor is connected to the MIQ/IDS module
(see section 3.7).
1.
Install the sensor in a flow through armature and provide the sample flow.
or
Submerse the sensor in the sample.
5.2

Checking

Prepare the test solution.
The sensor is in the sample.
Step 1:
The procedure is started from the measured value display with
<C> (see IQ S
During the check, the sensor is in maintenance condition.
All linked outputs remain in their current status.
CCESSORIES
system is in operation
ET
system operating manual).
ET
N
system operating manual).
ENSOR
ET
MIQ/IDS
N
.
ENSOR
ET
ba77277d01
01/2024