Check And Calibration; General Information; Function Check - Xylem wtw MIQ/IDS Operating Manual

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

Dissolved oxygen
Factory calibration
When does a check
or calibration make
sense?
Principle
Procedure
28
6.2

Check and calibration

6.2.1 General information

The check/calibration can be carried out on the following devices:
 On the IQ S
(see section 6.2.2 F
and section 6.2.3 C
 on an IDS meter
(see operating manual of the meter)
Details of the check or calibration may vary depending on the device
selected.
®
The FDO
925 is factory calibrated. The measuring characteristics of the sensor
cap remain stable for the specified service life. Thus, calibration is usually not
required.
A check or calibration can be useful in the following special cases:
 If the measured values appear to be implausible and it is assumed that the
service life of the sensor cap is over
 Routinely within the framework of the company quality assurance
A check/calibration is carried out in water vapor-saturated air (test and storage
®
vessel FDO
Check).
With the calibration procedure, the relative slope of the sensor is determined.
The calibration is evaluated based on the relative slope and the intensity
(successful <-> unsuccessful).

6.2.2 Function check

Generally, a check on the IQ S
specific details, please refer to the IQ S
Prepare the check and storage vessel FDO
The sponge in the FDO
moist (not wet).
N
ENSOR
ET
UNCTION CHECK
ALIBRATION
)
N
is carried out as follows. For system
ENSOR
ET
ENSOR
®
Check check and storage vessel must be
,
N
system operating manual.
ET
®
Check.
ba77277d01
MIQ/IDS
01/2024

Advertisement

Table of Contents
loading

Table of Contents