Turbidity; Measuring - Xylem wtw MIQ/IDS Operating Manual

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

Turbidity

Prerequisites
Preparatory activi-
ties
46
9
Turbidity
The MIQ/IDS module connects IDS sensors to the IQ S
Suitable sensors: see section 9.5 A
9.1

Measuring

CAUTION
Contact with the sample can lead to danger to the user!
Depending on the type of sample, suitable protective mea-
sures must be taken (protective clothing, protective goggles,
etc.).
 The IQ S
N
ENSOR
(see IQ S
N
ENSOR
 The sensor is connected to the MIQ/IDS module
(see section 3.7).
1.
Install the sensor in a flow through armature and provide the sample flow.
or
For measurement and calibration in containers:
Prepare the sensor for measurement (see P
46).
Perform the following preparatory activities when you want to measure:
 Avoid gas bubbles (e.g. air bubbles) in the test sample.
 Use suitable vessels for measurement and calibration
(see operating manual of the VisoTurb
 Heed the minimum depth of immersion for the sensor
1.
Fill the test sample into a lightproof measuring beaker up to a level of at
least 6 cm.
2.
When immersing the sensor in the test sample, hold the sensor at an
angle.
3.
For measuring, position the sensor upright.
CCESSORIES
system is in operation
ET
system operating manual).
ET
ENSOR
REPARATORY ACTIVITIES
®
900-P sensor).