Xylem wtw MIQ/IDS Operating Manual page 18

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

pH
18
Display
* Rinse the sensor.
* Immerse the sensor in the first
buffer solution.
* Wait for a stable measured
value.
* Enter the pH value of the first
buffer solution.
* Rinse the sensor.
* Immerse the sensor in the
second buffer solution.
* Wait for a stable measured
value.
* Enter the pH value of the sec
ond buffer solution.
Successfully calibrated.
End of the CAL_CON_2P cali
bration.
4.2.5 Calibrating with
Display indications during 1-point calibration with CAL CON 1P
Display
* Have any buffer solution
ready.
* Rinse the sensor.
* Immerse the sensor in the buf
fer solution.
* Wait for a stable measured
value.
Explanation
Follow the instructions on the display. As
soon as the measured value is stable, the
next display indication appears.
Enter the nominal pH value of the first buffer
solution according to the displayed tempera-
ture with
<
.
<OK>
Follow the instructions on the display. As
soon as the measured value is stable, the
next display indication appears.
Enter the nominal pH value of the second
-
buffer solution according to the displayed
temperature with
with
<OK>
The values determined for
metry potential
-
Calibration is finished.
Confirm with
The display returns to the measured value
display.
CAL CON 1P
Explanation
For this you can use any one buffer solution
if its pH value at the current temperature is
known. The calibration will be the more
exact the nearer the pH value of the buffer
solution is to that of the test sample.
Follow the instructions on the display. As
soon as the measured value is stable, the
-
next display indication appears.
and confirm with
>

and confirm
<
>

.
Slope
are displayed.
.
<OK>
ba77277d01
MIQ/IDS
Asym
and
-