Calibration Procedure On The Iq Sensor Net - Xylem wtw MIQ/IDS Operating Manual

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

MIQ/IDS
ba77277d01
8.2.2 Calibration procedure on the IQ S
The sensor is in the sample.
Step 1:
The procedure is started from the measured value display with
<C>
During calibration, the sensor is in maintenance condition.
All linked outputs remain in their current status.
Calibration:
Step 2:
Follow the instructions on the display.
Display
Reference voltage is deter-
mined. Take a sample now.
As soon as you have determined
the chlorine concentration of the
sample, you can continue with
the entry of the determined con-
centration. To do this, please
restart the calibration.
How to proceed
- Resume cal.
- Start new cal.
- Cancel cal.
If Resume cal. is selected:
Please enter the concentration
of sample.
Calibration done.
01/2024
(see IQ S
N
ENSOR
ET
N
ENSOR
ET
system operating manual).
Explanation
Take a sample as close to the
location and as soon as possi-
ble.
Determine the chlorine con-
centration of the sample taken
using a reference method.
Measurement can be contin-
ued during the determination.
Continue calibration by press-
ing
again.
<C>
Select the function
cal.
to continue entering the
reference value.
Enter the concentration of the
sample with
Confirm with
Calibration is finished.
Confirm with
The display returns to the
measured value display.
Chlorine
Resume
.
<
>

.
<OK>
.
<OK>
43