Informative Messages; Accessories - Xylem wtw MIQ/IDS Operating Manual

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

MIQ/IDS
IDS sensors
ba77277d01
Message code
EC9339

6.4.2 Informative messages

Message code
IC1339
IC3339
IC4339
IC5339
IC6339
IC7339
IS1339
6.5

Accessories

Description
Optical IDS oxygen sensor
Replacement sensor cap
01/2024
Message text
Calibration error, measurement disabled
Cause: Sensor cap is missing, leaky, depleted, or defective
* Clean sensor and space between cap and sensor accord
ing to op.instructions
* Screw on sensor cap securely
* Repeat calibration
* Replace sensor cap
Message text
Sensor has been successfully calibrated
* For calibration data, see calibration history
Factory calibration has been activated. Make sure the sen-
sor operates correctly.
Last valid calibration has been activated. Make sure the
sensor operates correctly.
Invalid user calibration has been replaced by last valid user
calibration. Caution! Wrong measured values possible.
Carry out a new successful calibration to make sure the
sensor operates correctly.
The invalid user calibration was replaced by the factory cal-
ibration. Caution! Wrong measured values possible. Check
whether the sensor operates correctly or carry out a new
successful calibration.
This message shows the result of the last check (see
section 6.2.2)
This message shows the information stored in the sensor
cap.
Model
®
FDO
925
®
FDO
925-P
SC-FDO® 925
Dissolved oxygen
-
Order no.
201300
201306
201310
37