MIQ/IDS
Menu item
Cell constant
(Sensor LR 925/01)
Cell constant
(Sensor TetraCon 925)
Temp. adjustment
Save and quit
Quit
IDS sensors
Standard for
calibration
ba77277d01
Selection/Values
0.090 ... 0.110 cm
0.450 ... 0.500 cm
-1.5 °C ... +1.5 °C
or
-2.7 °F ... +2.7 °F
-
-
7.4

Messages

IDS conductivity sensors do not generate any messages.
7.5

Accessories

Description
IDS conductivity measuring cell
Ultrapure water conductivity measuring cell
IDS connecting cable, 1.5 m
Description
KCL standard solution 0.01 M KCl, 6x 50
ml
For more accessories, refer to the operating manual of the sensor.
01/2024
Explanations
-1
Here you can set the cell constant if this
is necessary for a special application,
e.g. when using a flow-through arma-
-1
ture.
Here you can balance the temperature
sensor in the sensor against a reference
temperature measurement.
The settings are stored.
The display switches to the next higher
level.
The settings are not stored.
The display switches to the next higher
level.
Model
TetraCon
TetraCon
LR 925/01
LR 925/01-P
AS/IDS-1,5
Model
E-SET
Conductivity
Order no.
®
925
301710
®
925-P
301716
301720
301722
903850
Order no.
300572