Miq/Ids - Xylem wtw MIQ/IDS Operating Manual

Iq sensor net module with ids interfaces
Table of Contents

Advertisement

Technical data
Electrical data
Interfaces
Terminal
connections
56

11.2 MIQ/IDS

Nominal voltage
Power consumption
MIQ/IDS1
MIQ/IDS2
MIQ/IDS4
Protective class
Overvoltage
category
MIQ/IDS1
MIQ/IDS2
MIQ/IDS4
IQ S
N
con-
ENSOR
ET
nections
Terminal type
Terminal ranges
Cable cross-section of
mains voltage leading
cables
Cable feeds
max. 24 VDC via the IQ S
IQ S
N
system operating manual, chapter
ENSOR
ET
T
)
ECHNICAL DATA
1.5 W (with connected IDS sensor)
1.5 W (with 2 connected IDS sensors)
1.7 W (with 4 connected IDS sensors)
II
II
1 IDS interface
2 SENSORNET terminal connections
2 IDS interfaces
2 SENSORNET terminal connections
4 IDS interfaces
0 SENSORNET terminal connections
2 (MIQ/IDS1 and MIQ/IDS2)
0 (MIQ/IDS4)
Additional switchable SENSORNET terminator (ter-
minating resistor)
Screw-type terminal strip, accessible by opening the
lid
Solid wires:0.2 ... 4.0 mm
AWG 24 ... 12
Flexible wires:0.2 ... 2.5 mm
Europe:1.5 ... 4.0 mm
USA:AWG 14 ... 12
4 cable glands M16 x 1.5 on the underside of the
module
N
(for details see
ENSOR
ET
2
2
2
ba77277d01
MIQ/IDS
01/2024

Advertisement

Table of Contents
loading

Table of Contents