IGEA OSTEOSPINE User Manual page 27

Table of Contents

Advertisement

Available languages

Available languages

IGEA
IGEA
-
Do not use the device in the presence of inflammable gases or rooms in which an explosion risk
exists.
-
Do not connect any part of unit to other equipment or devices.
-
Do not connect to OSTESPINE any part not intended for use and not supplied by the
Manufacturer.
-
Do not use the device in the proximity of short wave or microwave treatment appliances, as it
may alter the output signals.
-
Never handle any components of the system with wet hands, in particular do not connect the
external battery charger to the mains power supply.
-
Do not use the device in areas or situations where it might get wet. The device remains secure
even in the event of penetration of liquids within the container, but may be damaged and no
longer be able to deliver therapy.
-
Never immerse any of the component parts of the system in water or liquids of any type and
never pour any liquids over the same. In the event of the accidental immersion of the generator
in liquids it must no longer be used and must be returned to an authorized assistance centre
or manufacturer for checking/repairs.
-
Do not expose the system to heat and do not dispose of in fire, danger of explosion!
-
Before each treatment session always ensure that cables connecting the generator to the
electrodes or battery charger cable are not damaged; should they be damaged contact an
authorized assistance centre.
-
Before each treatment session always ensure that electrodes are in good conditions; should they
be damaged contact an authorized assistance centre.
-
Avoid excessive pulling or stretching of the connection cable which comes out the electrode and
which makes contact with the conductive material of the same, as any cable damage will render
the electrode unfit for use.
-
Avoid any mechanical shocks or blows to the appliance and its components during transportation
and handling.
-
The rechargeable battery is classified as polluting substances; it must be disposed of in
accordance with the local governmental legislation about pollution.
-
The device is equipped with self-control mechanisms of the correct functioning; any anomaly that
may occur is indicated by the system and is described in the instruction manual. Any malfunctions
therefore do not affect therapeutic efficacy.
-
IGEA recommended a routine maintenance procedure for the operating parameters check at
regular intervals not exceeding 24 months of use to ensure reliable performance; ccontact IGEA
or an authorized assistance centre.
-
The generator can be cleaned with a damp cloth and neutral detergents; before cleaning the
generator ensure that the device is switched off and disconnected from the battery charger.
5.2
Contraindications and adverse effects
The numerous international studies carried out have not shown there to be any adverse effects as the
result of the treatment as a whole, however these precautions are necessary:
-
In the event of either suspected or confirmed pregnancy before using the device inform the doctor
who has prescribed the therapy, who will decide on whether to continue with the specific
treatment or not.
-
Those fitted with pace-makers are advised to check the features of their pace-maker device with
their cardiologist; and in any case avoid any treatments which involve the application of
electrodes in the vicinity of the pace-maker itself;
-
In the event of any skin damage or irritation in the zone in which the electrodes are to be
positioned, it is necessary to consult the doctor who prescribed the therapy for advice on
alternative positioning of the electrodes or the advisability of suspending the actual treatment.
-
Therapy should not be performed concurrently with other therapies or applications of medical
devices that release energy to the patient body through the use of electrodes, in particular if they
OSTEOSPINE – User Manual
OSTEOSPINE – User Manual EN
25
25
25