IGEA OSTEOSPINE User Manual page 23

Table of Contents

Advertisement

Available languages

Available languages

IGEA
IGEA
9. The system is provided with a number of electrodes corresponding to the average demand for
treatment; to receive more electrodes contact the manufacturer or an authorized service centre.
10. Do not to re-use the electrodes in the event that the adhesive potential is visibly diminished; the
electrodes must firmly and thoroughly adhere to the skin.
11. When possible it is advisable the use of a cotton or rubber belt or any other suitable mean in order
to allow the electrodes to firmly and thoroughly adhere to the skin.
12. The electrodes and the parts of the device that may come into contact with the skin usually do not
cause any allergic reaction. in case of irritation in the area of use please consult your doctor.
13. The electrodes package shows the expiring date of the same. Do not use electrodes after the
expiration date printed on the package.
3.4
System conditions indications
During operation different situations may occur, indicated by the combination of indicators and buzzer
described below.
STIMULATOR CONDITION
Stimulator on and searching for signal, the
electrodes neither yet applied to the patient
nor connected to the generator. When
switch on the generator it takes a few seconds
to check that the electrodes are present and
applied on the body, then the treatment signal
starts.
Stimulator on with electrodes connected
and in correct operational stage.
Low Battery.
The stimulator is on and works correctly, but
the battery is almost flat, from the signal alarm
the appliance has got some working hours.
Flat Battery.
The stimulator is on and in operation but the
battery is totally flat and can no longer power
the circuit. The battery needs charging in order
to resume therapy.
Battery recharging in progress.
The stimulator is connected to the external
battery charger and the battery is charging.
Completed charging operation, should the
connector not be disconnected from the
battery charger, the red indicator remains on,
otherwise the red light goes out whenever the
connector is disconnected from the battery
charger.
Short-circuited electrodes or cables.
The stimulator is on but there is a functional
anomaly due to a direct contact of the
electrodes or to a short-circuiting of the
cables. Check that the electrodes or their
connecting cables are not in direct contact with
each other.
OSTEOSPINE – User Manual
INDICATORS AND BUZZER
Yellow indicator Continuous light on
Red indicator
Buzzer
Yellow indicator On and flashing
Red indicator
Buzzer
Yellow indicator On and flashing
Red indicator
Buzzer
Yellow indicator On and flashing
Red indicator
Buzzer
Yellow indicator Off
Red indicator
Buzzer
Yellow indicator Continuous light on
Red indicator
Buzzer
OSTEOSPINE – User Manual EN
Off
Emits an intermittent sound
until the problem is resolved
Off
Off
On and flashing
Buzzes for a second and then
switches off
Continuous light on
Emits a "beep" every 5 sec for
a total of 30 seconds after which
the generator switches off.
On and flashing for the whole
recharging time. At the end of
charging,
continuous
remains on.
Off
Off
Off
light
21
21 21