Running The Program; Setting The Specification - Chroma 63200 Series Operation & Programming Manual

High power dc electronic load
Hide thumbs Also See for 63200 Series:
Table of Contents

Advertisement

High Power DC Electronic Load 63200 Series Operation & Programming Manual

4.2.3 Running the Program

Press LOAD ON/OFF to run program when the program function is selected. The display
shows as below.
The upper line shows the executed program and sequence number while the lower line shows
the Load, key, test result and short status.
ON/OFF:
It shows the Load input status.
KEY:
It displays when MANUAL mode is active and waiting for key input.
PASS/FAIL:
It shows the test result compared with SPEC setting.
S:
It display if Load is Short now.
When executing the program, the setting of sequence will recall files from EEPROM, and the
SPEC function is always ON. All function keys are disabled until LOAD ON/OFF is pressed
to stop the program execution, or the program run finishes. When the program run stops or
finishes, the LCD will display as follows.
It means that all sequences have passed in the program tests. If the test fails, LCD will show
as follows.
PROG. XX indicates the file number of the program failed, 1 to 10. In addition, 1, 2, 3...10
shown by LCD stand for the failed sequence numbers. During program chain testing, if the
failed program files are more than one set, you can use ▲, ▼ to read the contents of failed
programs.

4.2.4 Setting the Specification

The GO/NG key is to enable/disable specification checking function. The Load will compare
measurement data with the set specifications for HIGH and LOW boundary when the SPEC
TEST is ON, and the LED, GO/NG, is lighted on the panel. To set specifications for the load;
please refer to the section 3.5 at the end of each mode for parameters setting. The
specification unit for CC, CR, and CP modes is volt while CV mode is current. There are
three levels for each mode: CENTER, HIGH and LOW. The CENTER level must be set by
the value of channel input reference level. The HIGH and LOW levels can be set by the
value or percentage selected in configuration SPEC.ENTRY MODE. The HIGH/LOW
percentage range is from 0 to 100%.
The default setting of HIGH and LOW is 100%. The CENTER value is half of the range.
Please refer to section 4.2.5 for the specification selection set by Value or Percentage.
4-12
PROG. 1
SEQ. 1
[ON ][ KEY ] [ PASS ] [ S ]
PROGRAM OFF
RESULT :
PASS
PROG. XX
Err. REPORT
1, 2, 3,.......10