Status Sub-System - Chroma 63200 Series Operation & Programming Manual

High power dc electronic load
Hide thumbs Also See for 63200 Series:
Table of Contents

Advertisement

High Power DC Electronic Load 63200 Series Operation & Programming Manual

7.2.12 STATus Sub-system

STATus
:QUEStionable
:CONDition
:ENABle
[:EVENt]
:NTRansition
:PTRansition
STATus:QUEStionable:CONDition
Description
Query Syntax
Return Parameter
Query Example
Return Example
STATus:QUEStionable:ENABle
Description
Syntax
Parameter
Bit Configuration of Questionable Status Register
Bit Position 15 14 13 12 11 10 9
Condition
0
Bit Weight
Example
Query Syntax
Return Parameter
Query Example
Return Example
STATus:QUEStionable[:EVENt]?
Description
Query Syntax
Return Parameter
Query Example
Return Example
STATus:QUEStionable:PTRansition/NTRansition
Description
Syntax
Parameter
Example
7-28
Real-time ("live") recording for Questionable data
STATus:QUEStionable:CONDition?
<NR1>
STAT:QUES:COND?
6
Mask the bits in the Event register that are allowed to be summed
into the QUES bit of the Status Byte register.
STATus:QUEStionable:ENABle
0
0
0
0
0
0
STAT:QUES:ENAB 24
STATus:QUEStionable:ENABle?
<NR1>
STAT:QUES:ENAB
24
Record all Questionable conditions that have occurred since last
time the register was read.
STATus:QUEStionable:EVENt?
<NR1>
STAT:QUES:EVEN?
24
Programmable filters determine what type of transition (0-to-1 or
1-to-0) in the Condition register will set the corresponding bit of the
Event register.
STATus:QUEStionable:PTRansition <NRf>
STATus:QUEStionable:NTRansition <NRf>
0 ~ 65535
STAT:QUES:PTR 4
Return the channel status.
8
7
6
5
4
0
0
0
0 OT RV OP OV OC
16
Return the setting of the Status
Questionable Enable register.
Return the contents of the Questionable
Event register.
Set OP(over power bit 2) to 0-to-1.
3
2
1
0
8
4
2
1

Advertisement

Table of Contents
loading

This manual is also suitable for:

63211632212