Figure 3-13 Load Surge Capability (Dynamic Mode)
Note
1.
The Initial state under Static Mode should last at least 1 second.
2.
This load surge capability will be regulated by the temperature de-rating
characteristics. (Refer to
3.
Examples below assume the use of the Model 63201 load with a continuous
rating of 2600W/300A/1-80VDC.
4.
63211 and 63212 have no surge capability.
Example 1: STATIC LOADING
The Model 63201 can be overdriven to approximately 5200W (200% of its rated continuous
power rating) for 0.06 seconds when the starting power is 650W (25% of its rated power).
This is represented by DOT on the blue curve in Figure 3-12.
Figure 3-12 Load Surge Capability (Static Mode)
Note 1
in Specifications)
Operation Overview
3-11