Usage

EN-
Usage
US
9

Operation

CAUTION
If the sensor is damaged, harmful
substances may exit from the unit
Check the sensor and cable for dam-
age before every use.
Do not use the sensor if it is damaged.
Use a right-angle holder.
NOTICE
Damage to the sensor
The interior parts of the sensor can be
damaged by mechanical forces.
Do not drop the sensor.
Do not subject the sensor to any pres-
sure.
Do not fold, clamp or crush the cable.
Do not carry the sensor by the cable.
We highly recommend using the sensor with a
right-angle holder. This prevents positioning
errors and protects the sensor against mechani-
cal damage (e.g. through an alarmed patient bit-
ing the sensor).
Operating the sensor with DBSWIN or via
VistaEasy, you can generate X-ray images in two
acquisition modes:
– Standard (26.3 LP/mm)
– Low resolution (13.15 LP/mm)
The transmission time is just a few seconds
depending on sensor size and settings.
24
9.1

Taking an X-ray image

Check before every use:
ü The sensor and cable are not damaged.
ü The sensor and cable have been cleaned and
disinfected.
Wear hand protection.
Start the computer.
Start the imaging software.
Select the desired image mode.
Make SensorX ready for image acquisition in
the imaging software.
WARNING
Danger of cross contamination when
not using the hygienic protective
cover or when using the hygienic pro-
tective cover more than once
The materials used for the hygienic pro-
tective cover are not suitable for reproc-
essing. This can result in malfunctions,
leaks and material failure.
Do not use the unit without the
hygienic protective cover.
Do not re-use the hygienic protective
cover (disposable item).
Push the hygienic protective cover (see "3.2
Accessories") over the sensor.
2121100020L29 2101V007