Chapter 4
Room Planning
Section 1.0

Required Systems Clearances

Consult your local GE Sales and Service Representative about your specific needs.
Some possible room size dimensions are shown in the tables below.
Recommended Scan Room
Size
3962 x 6299 mm
(13ft. x 20ft. 8in. )
Table 4-1 List of Scan Room Layouts
Limited Access Rooms (width only):
Recommended Ctrl Room Size
(Standard Console Table)
3962 x 2743 mm
(13ft. x 9ft. )
Table 4-2 List of Control Room Layouts
Short Footprint (length only):
Additional component dimensions are available in
Consult your local General Electric Project Manager of Installation for your appropriate room
specifications.
For equipment clearance requirements, refer to
Service clearances must be maintained around equipment for full operation, service and safety.
Cable length is an important consideration in room layout. A set of short cables (B71172CA) and a
set of long cables (B71182CA) are optional for BrightSpeed system.
Note, also, that where possible, the cables should enter the gantry from the rear, utilizing the rear
cable cover assembly. Alternate cable entry is possible at the center of the gantry (refer to the
Installation template).
Excess cable length cannot be stored behind the console or PDU.
Long cable must not be cut or shortened.
Excess cable may be stored in conduits, a cable storage box if present, or the floor duct,
provided sufficient space is available. Observe the fill rate for each option. If there are
questions regarding local electrical or building codes, consult the project electrical contractor
or electrician.
All NEC 70-E Electrical Regulations regarding conduit or duct fill must be observed.
14
Minimum Width: 3353 mm (11ft. ); 356 mm (14in. ) cover to wall
Average Width: 3708 mm (12ft. 2in. ); 711 mm (28in.) cover to wall
Minimum length: 5480 mm (18ft. )
Chapter 4 - Room Planning
B
S
RIGHT
PEED
Typical Scan Room Size
3912 x 5868 mm
(12ft. 10in. x 19ft. 3in.)
Recommended Ctrl Room
Size (Freedom Workspace
Console Table)
3861 x 2500 mm
(12ft. 8in. x 8ft. 2in. )
Figure 4-7
through
Section
2.0. Remember, sufficient Regulatory and
E
, E
, E
: P
-I
LITE
DGE
XCEL
RE
NSTALLATION
Clearance
Requirements
see
Figure 4-1
Figure 4-15
of this document.
Page 51