Electrical Requirements
Figure 5-1
WARNING
THE REMOTE SPEAKER KIT SHALL NOT BE INSTALLED IN THE PATIENT VICINITY.
Patient Vicinity Definition : see Patient Environment Equipment in
2.2.3 Room Layout Considerations on
page 80

5.9 MAC-LAB System EX

For location of MAC–LAB Acquisition Unit Floor mount kit 408431–001, refer to Electrical Ducts in
5.3.2 Cable Channeling on page
156.
For MAC–LAB System EX, refer to:
- Marquette document P/N 2000465–001 Preinstallation Guide

5.10 Advantage Windows

A power cord is supplied with Advantage Windows. It is connected to outlet fed by the main disconnect
room device.
Verify that the ground between AW, AW spliter and monitors ground are equipotential (0.1 ohm
equipotentiallity between room ground point (busbar) and wall outlets).
176