Transducer
Thermocouple,
MX Modules
4.4 MX1609KB-R Thermocouple Module
Up to 16 type K thermocouples (NiCrNi) can be connected to the MX1609KB-R
module for measuring temperatures.
K-type
Status LEDs
The following table gives the descriptions for all LED states.
System LED
Green
Red
Orange
Orange flashing
--
--
--
--
A4089-5.0 en HBM:Public.
MX1609B-R
●
Channel
Description
LED
--
System is error free
--
System error
Orange
System is not ready; boot procedure is
running
Orange
System is not ready; download is active
flashing
Green
Connection is error free
Green
TEDS data is valid
flashing
(LED flashes for 5 s)
Orange
Transducer identification/sensor scaling
is running
Red
No sensor connected
Channel error (incorrectly parametrized,
connection error, invalid TEDS data)
Overload of sensor supply
Wiring diagram
72