Transducer
Voltage,
60 V, 10 V, 100 mV
Current-fed piezoelectric transducer
(IEPE, ICP®)
Current,
20 mA
Use ODU 14-pin to BNC adapter 1-KAB430-0.3.
MX Modules

4.3 MX1601B-R Standard Module

Up to 16 configurable inputs for DC voltage sources (60 V, 10 V, 100 mV), DC
current sources (20 mA) or current-fed piezoelectric transducers (IEPE, ICP®) can
be connected to the MX1601B-R module.
All measuring channels are electrically isolated from one another and from the power.
When using transducer excitation, electrical isolation from the supply voltage of the
amplifier is rescinded.
Status LED
The following table gives the descriptions for all LED states.
System LED
Green
Red
Orange
Orange flashing
A4089-5.0 en HBM:Public.
MX1601B-R
    ●   
Description
System is error free
System error
System is not ready; boot procedure is running
System is not ready; download is active
Wiring diagram