Status Register - Rohde & Schwarz R&S FSW-K106 User Manual

Lte nb-iot measurement application (downlink)
Table of Contents

Advertisement

®
R&S
FSW-K106
The ASCII character # introduces the data block. The next number indicates how many
of the following digits describe the length of the data block. The data bytes follow. Dur-
ing the transmission of these data bytes, all end or other control signs are ignored until
all bytes are transmitted. #0 specifies a data block of indefinite length. The use of the
indefinite format requires an NL^END message to terminate the data block. This format
is useful when the length of the transmission is not known or if speed or other consid-
erations prevent segmentation of the data into blocks of definite length.

6.3 Status register

The LTE NB-IoT measurement application uses the standard status registers of the
FSW (depending on the measurement type). However, some registers are used differ-
ently. Only those differences are described in the following sections.
For details on the common FSW status registers refer to the description of remote con-
trol basics in the FSW user manual.
*RST does not influence the status registers.
STATus:QUEStionable:SYNC register
The STATus:QUEStionable:SYNC register contains application-specific information.
If any errors occur in this register, the status bit #11 in the STATus:QUEStionable
register is set to 1.
Each active channel uses a separate STATus:QUEStionable:SYNC register. Thus, if
the status bit #11 in the STATus:QUEStionable register indicates an error, the error
may have occurred in any of the channel-specific STATus:QUEStionable:SYNC reg-
isters. In this case, you must check the register of each channel to determine which
channel caused the error. By default, querying the status of a register always returns
the result for the currently selected channel. However, you can specify any other chan-
nel name as a query parameter.
Table 6-2: Meaning of the bits used in the STATus:QUEStionable:SYNC register
Bit No.
0
1
2 to 5
6
7
8
User Manual 1178.5937.02 ─ 16
Meaning
Configured frame not found
Sync not found
Unused
Auto level no signal
Setting mismatch
Signal analysis error
Remote control
Status register
79