Magnetic Field:
Maximum number of stimulations:
Pulse Characteristics:

10.5 REAR PANEL

10.6 GENERAL

The Magstim Company Limited 3001-23-04
The magnetic field produced by the Magstim 200
type of coil connected. Full details of the Magstim 2nd
Generation Coils, and Remote Control Coils, are given in their
operating manuals, P/N 3100-23 and 3179-23, respectively.
The maximum number of stimulations that can be achieved with a
2
Magstim 200
at an ambient temperature of 20
between discharges of 4 seconds, before the coil exceeds its
permitted operating temperature, depends on the type of coil
connected. Details of the maximum number of stimulations
possible for each of the Magstim 2nd Generation, and Remote
Control, Coils are given in their operating manuals, P/N 3100-23
and 3179-23, respectively.
100µs approximate rise time, 1ms duration
ISOLATED TRIGGER I/O Port (see Section 4).
The maximum non-destructive voltage range that can be applied to the
external connections with respect to ground is: - 0.3 – 5.3V
Connection
Trigger Input
Trigger Output
Size:
Width
Height
Depth
Weight
20.4 kg
o
Logic Low
<0.8 V
<0.8 V
460 mm (18.1 inches)
160 mm (6.3 inches)
375 mm (14.8 inches)
28
2
depends on the
C, with an interval
Logic High
>4V
>4V