2
can be put into the ARMED mode by momentarily
The Magstim 200
pushing the RUN button. This can only be achieved if the stimulating coil
is connected to the COIL OUTPUT socket (18) and the REPLACE
COIL symbol (9) is not illuminated. In the ARMED mode the ARMED
symbol (11) will illuminate and the instrument will charge to the level
displayed on the OUTPUT % DISPLAY (4).
2
The Magstim 200
can be triggered, and a magnetic pulse produced, by
pushing the TRIGGER button when the READY symbol (12) is
illuminated and one of the safety buttons at the base of the stimulating
coil is activated. The instrument will continuously charge and discharge if
the TRIGGER button is pressed and held.
The output level setting as shown on the OUTPUT % DISPLAY (4) can
be increased, or decreased, by rotating the control clockwise or anti-
clockwise respectively. Each complete 360º rotation changes the output
level by approximately 20%.
2
The Magstim 200
can be put into the safe inactive default mode by
momentarily pushing the STOP button. In this mode the DEFAULT
CONDITION indicator (10) will illuminate and the instrument will
discharge internally.
Note: The default mode is automatically selected if the coil is
disconnected, becomes faulty, the surface temperature of the connected
Coil has risen above its set limit, the SYSTEM STOP button is activated,
the instrument detects an internal fault, or if the READY symbol has
been illuminated for over 1 minute without the instrument having been
triggered.
This socket is used to connect the stimulating coil to the Magstim 200
Note that the unit cannot be armed and triggered unless the stimulating
coil is connected to this output. Ensure that the locating spot on the coil
plug is correctly lined up with the coil output socket before inserting the
plug. When the coil plug is fully engaged in the socket, lock the plug in
place by turning the black locking ring clockwise.