2
Magstim 200
SECTION 2:
CAUTION: Magnetic Stimulators have not been cleared by the FDA for
cortical stimulation. Investigational human uses require that the user complies
with the FDA regulations regarding Investigational Devices. Please visit
USA Only
www.fda.gov/cdrh or www.magstim-us.com for more information.
CAUTION: In the USA, Federal Law restricts this device to sale by or on
the order of a practitioner licensed by the law of the State in which he/ she
practices to use or order the use of the device.
Attention: Consult accompanying documentation before using the Magstim
2
200
CAUTION: The Magstim 200
or in the vicinity of, patients or subjects with cardiac demand pacemakers,
implanted defibrillators, or other electronic implants.
The Magstim 200
pulses. The induced eddy current is of sufficient magnitude to stimulate nerves
and muscle.
The strong magnetic pulses generated by stimulating coils induce eddy currents
in any conductive medium such as the human body, nearby metallic objects or
electronic devices.
NB Particular care must be taken to ensure that leads connected directly to the
patient, or other equipment, are not in a position where the stimulating coil can
couple resulting in currents being induced in them.
Do not discharge the Magstim with the stimulating coil in the vicinity of
metallic objects or these may be projected, moved and/or damaged.
The Magstim 200
presence of flammable anaesthetics.
When the magnetic pulse is delivered, a discharge click is produced by the
Magstim 200
Where possible avoid the use of the stimulating coil near the ears. The use of
ear plugs is recommended.
High voltages are present within this instrument. Do not remove covers. Refer
servicing to qualified personnel.
The Magstim Company Limited 3001-23-04

WARNINGS AND PRECAUTIONS

.
2
, its coils and accessories generate high intensity magnetic
2
must not be used in an explosive atmosphere or in the
2
and its stimulating coil. This discharge click may startle.
2
and its stimulating coils must not be used on,
3